143
Input/Output/Measurement Specications
Measurement time
Settings of contact check function and comparator function ON OFF
Contact-checking time 2.3 ms 0.0 ms
Comparator measurement time 0.2 ms 0.0 ms
Setting of
measurement speed
50 Hz 60 Hz
Internal integration
time
FAST 4.1 ms 4.1 ms 2 ms
FAST2 13.7 ms 12.7 ms 0.5 PLC
MED 23.7 ms 20.7 ms 1 PLC
SLOW 109 ms 93 ms 4 PLC
SLOW2 320 ms 320 ms 13 PLC
• Using the external control (EXT I/O) extends the INDEX and the EOM times as follows:
INDEX time (Contact check delay time) + (Contact-checking time) + (Delay time) + (Measurement time)
EOM time INDEX + (Comparator measurement time) + 0.4 ms
To obtain the calculated resistance value based on the measured voltage, add 1.0 ms.
• Self-calibration time: 5 s or less
Measurement time examples (With the measurement speed set to FAST)
Contact check
Comparator
measurement
Power frequency:
50 Hz
Power frequency:
60 Hz
INDEX EOM INDEX EOM
OFF OFF 4.1 ms 4.5 ms 4.1 ms 4.5 ms
OFF ON 4.1 ms 4.7 ms 4.1 ms 4.7 ms
ON OFF 6.4 ms 6.8 ms 6.4 ms 6.8 ms
ON ON 6.4 ms 7.0 ms 6.4 ms 7.0 ms
Temperature
measurement
accuracy
Display range -40.00°C to 80.00°C
Displays [--.--°C] with a temperature sensor not connected.
Accuracy range -40.00°C to 80.00°C
Measurement accuracy See “Specications of Model Z2011 Humidity Sensor” (p. 144).
Measurement period 2 s ±0.2 s
Humidity
measurement
accuracy
Display range 0.0% RH to 90.0% RH
Displays [--.-%rh] with a humidity sensor not connected.
Accuracy range 20.0% RH to 80.0% RH
Measurement accuracy See “Specications of Model Z2011 Humidity Sensor” (p. 144).
Measurement period 2 s ±0.2 s
D/A output accuracy Output accuracy (Current measurement accuracy) ±0.2% f.s.
Temperature coecient ±0.02% f.s./°C
Response time (Measurement time) + 1 ms at a maximum
11
Specications