EasyManua.ls Logo

Hioki Super megohm SM7110 - Page 155

Hioki Super megohm SM7110
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
149
Function specications
Interlock function Operation details Disables output of measurement voltage and measurement using
an external input signal.
When the setting is enabled, the interlock is deactivated by
inputting the On signal or connecting the terminals with each
other.
If the interlock function is activated during a measurement, the
measurement stops.
Measurement cannot be performed by either the keys or
communication in the interlocked state.
Setting
ON (enable)/OFF (disable)
Input terminal BNC (Internally in parallel with external I/O)
Sequence program
function
Operation details Sets the sequence of discharge, charge, measurement, and
discharge and executes it.
Setting
ON/OFF (If OFF is set, continuous measurement is executed.)
Save number
0 to 9
Allowable time setting range for each pattern (The setting unit can be switched between
milliseconds and seconds.)
Measurement:
1 ms to 999.9 s, default: 0.1 s
Other than measurement:
0 ms to 999.9 s
Resolution
In milliseconds: 1 ms
In seconds: 0.1 s
If the unit is changed from milliseconds to seconds, rounded-o󰀨
values are displayed.
Panel saving, panel
loading
Operation details Saves and loads the measurement conditions specied by the
panel number.
Numbers of panels 50
Panel name 10 characters (Alphabets or numbers)
Saved contents Measured value display mode, voltage output set value, setting of
voltage set value for resistance calculation, measurement speed,
range, trigger mode, delay, average, contact check, self-calibration
setting, resistivity measurement settings, comparator, judgment
sound, interlock, sequence, xture capacitance open correction
value
11
Specications

Table of Contents