150
Function specications
Voltage monitor
check
Operation details When a monitored voltage value falls within the specied range,
the V_CHECK_GO signal switches to the on state.
Setting
ON/OFF
Settable range
±2% to ±20% (Resolution: 1%)
Display
ON: [V.CHK] is displayed during normal operation.
[V.CHK] is displayed during erroneous operation,
highlighted in reverse video
OFF: Non-displayed
Voltage monitor range for measured voltage
SM7120
MES.V Voltage monitor range Display range of voltage
monitor
0.1 V to 9 V 10.000 V 10.000 V
9.1 V to 90 V 100.00 V 100.00 V
90.1 V to 900 V 1000.0 V 1000.0 V
900.1 V to 2000 V 2000.0 V 2000.0 V
SM7110
MES.V Voltage monitor range Display range of voltage
monitor
0.1 V to 9 V 10.000 V 10.000 V
9.1 V to 90 V 100.00 V 100.00 V
90.1V to 1000 V 1000.0 V 1000.0 V
DA output function Operation details Outputs a voltage corresponding to the measured current.
Even when a resistance value is displayed, outputs a voltage
corresponding to the measurement current.
Output voltage 0 V DC to 2 V DC (Outputs 2.0 V at the maximum display value of
the current range)
An error or Current Over Range (out-of-range) results in a voltage
of 2.5 V outputting.
Maximum output
voltage
5 V DC
Output impedance 1 k
Ω
Number of bits 12 bits
Output terminal BNC terminal
Double action
function
Operation details After the STOP key is pressed, the START key is enabled for only
1 second.
Setting
ON / OFF
Reset function Reset Resets the settings excluding the panel data to the factory default.
System reset Resets all the settings including the panel data to the factory
default.
Self-test function Self-test on start-up Executes the ROM/RAM check.
Error display See “Error display and solution” (p. 160).