64
Setting Pre-charge
4.8 Setting Pre-charge
The CHARGE terminal can be used to charge measured objects before measurement (pre-
charge). Thus, this setting enables required times for automatic measurements for objects such as
capacitors to decrease.
IMPORTANT
If the measurement voltage setting or the charge setting is changed after the current limiter has
been set and if the current limiter value exceeds the upper limit, the current limiter is set to the
largest value among the settable values.
See “4.1 Starting Measurement After Measured Value Becomes Stable (Delay Function)” (p. 49).
To connect objects to the instrument, see “2.6 If Measured Object Is to Be Charged Before
Measurement (Pre-charge)” (p. 33).
Procedure to display the settings screen: (Measurement screen) MENU key >
[MEAS] tab
[ON] Charges objects to be measured.
[OFF] Does not charge objects to be measured. (default setting)