Further Accelerating Measurement (Function of Updating Drawing During Measurement)
4
Set the electrode constant (K).
0.01 to 999.99 (Can be set in increments of 0.01, default setting: 500.00)
Settle
Cancel
Move to the next
digit (left or right)
Change the
numerical value
(up or down)
4.14 Further Accelerating Measurement (Function of
Updating Drawing During Measurement)
Usually, measured values displayed on the screen are updated every time a measurement
completes. In this case, you need to wait until the screen is updated for some time, a period from
when a measurement completes (the EOM is output) to when the next TRIG is accepted. Set this
function setting to [OFF] to accelerate further the measurement by omitting measured value display
updating.
Procedure to display the settings screen: (Measurement screen) MENU key >
[MEAS] tab
[ON] Updates the measured value display every measurement (default setting).
[OFF] Does not update the measured value display (Update the display when the setting is changed).