OPERATING INSTRUCTIONS
ThinPrep 2000 Processor Operator’s Manual
5A.19
5A
5. Dispose of the used ThinPrep Pap test filter cylinder using appropriate laboratory procedures.
A
ThinPrep Pap test filter must be used only once and cannot be reused.
6. Remove the PreservCyt sample vial from the instrument and recap it firmly. Be sure to line up the
torque line on the cap with the torque line on the vial. Refer to Figure 5A-13.
Figure 5A-13 Capping the PreservCyt Sample Vial
7. Do not discard the sample vial until it has been determined that no additional slides are needed.
Refer to Chapter 3, “PreservCyt Solution”, for information regarding solution disposal and sam-
ple storage.
WARNING: See Package Insert Before
SPecimen Collection.
NAME:
ID#:
3256723-46
Line on cap and
line on vial should
meet or slightly
overlap. If the cap
on the vial does
not have a line,
ensure the cap is
tightened
securely.