INSTRUMENT TROUBLESHOOTING
6.32
ThinPrep 2000 Processor Operator’s Manual
Waste System Failure
Reason for Message
This message appears when the instrument fails to detect its target negative pressure in the waste
bottle during idle mode or at the start of a sequence.
Possible Causes
Waste bottle cap is not secure
Waste fittings are disconnected from back of instrument
Waste tubing is disconnected or obstructed at any point
System hardware malfunction
Waste filter is wet