7
MAINTENANCE
7.4
ThinPrep 2000 Processor Operator’s Manual
FILTER CAP CLEANING
Clean the filter cap daily. It is important that the top surface of the filter cap and the cap seal O-ring
are free of debris for proper operation of the system.
Wipe the entire filter cap with a lint-free wipe dampened with deionized or distilled water to remove
deposits. Dry the filter cap after cleaning.
Note:
Handle the filter cap gently. Do not hit it against hard surfaces.
Figure 7-2 Filter Cap
SECTION
C
Cap seal O-ring
Filter seal O-rings