OPERATING INSTRUCTIONS FOR PROCESSING COBAS AMPLICOR™ CT/NG
SAMPLES
ThinPrep 2000 Processor Operator’s Manual
5B.5
OVERVIEW OF LOADING THE THINPREP
®
2000 PROCESSOR
The following sections describe in detail the methods for loading the ThinPrep 2000 processor. Before
initiating a run, the following supplies must be loaded into the processor in the order indicated.
• Fixative vial
• ThinPrep Pap test filter and filter cap (filter assembly)
• PreservCyt sample vial
• ThinPrep Slide
The figure below shows the ThinPrep 2000 processor after loading of the supplies is complete.
Figure 5B-2 ThinPrep 2000 Processor Loaded with Supplies
Slide handler/ejector
Fixative
vial
Fixative
bath holder
Cap seal
Bobbins
Filter assembly
PreservCyt
sample vial
Sample vial
holder
Slide