MAINTENANCE
ThinPrep 2000 Processor Operator’s Manual
7.15
7
Make sure the new waste filter is correctly oriented. The filter specification notations on the
waste filter must be on the connector side of the filter, not the waste bottle side.
7. Attach the tubing to the top of the new waste filter.
8. Reattach all the waste connectors to the rear of the instrument. Refer to Chapter 2, “ThinPrep
2000 Installation” for more information.
9. Turn on instrument power.
10. Run a blank (PreservCyt Solution vial with no cells) on the processor to test the operation of the
instrument.