EasyManua.ls Logo

Honeywell UDC2500 - Table 4-18 Set up Procedure

Honeywell UDC2500
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Monitoring and Operating the Controller
4/07 UDC2500 Universal Digital Controller Product Manual 111
Set up Procedure
Use the following procedure (Table 4-18) to:
select two sets,
set the switch-over value,
set tuning constant value for each set.
Table 4-18 Set Up Procedure
Step Operation Press Result
1
Select Control
Set-up Group
SetupSetup
Until you see:
Upper Display = SET
Lower Display = CONTRL
2
Select PID SETS
FunctionFunctionFunction
Until you see:
Upper Display = (available selections)
Lower Display = PID SETS
3
Select PID SETS
Function
or
To select the type of function. Available selections are:
ONE—1 set of constants
2 KBD—2 sets, keyboard selectable
2 PR—2 sets, auto switch at PV value
2 SP—2 sets, auto switch at SP value
4
Set Tuning
Values for Each
Set
FunctionFunctionFunction
Refer to “TUNING” Set up group, subsection
3.4 and set the
following tuning parameters:
PB or GAIN*
RATE T*
I MIN or I RPM*
CYCT1 or CTIX3*
PB2 or GAIN2**
RATE 2T**
I2MIN or I2RPM**
CYC2T2 or CT2X3**
*PIDSET1 will be used when PV or SP, whichever is
selected, is greater than the switchover value.
**PIDSET2 will be used when PV or SP, whichever is
selected, is less than the switchover value.
5
Set Switchover
Value for 2PR or
2SP Selection
FunctionFunctionFunction
Until you see:
Upper Display = (the switchover value)
Lower Display = SW VAL
or
To select the switchover value in the upper display.

Table of Contents

Other manuals for Honeywell UDC2500

Related product manuals