EasyManua.ls Logo

Honeywell UDC2500 - Table 6-4 Auxiliary Output Calibration Procedure (Numeric Code 50000)

Honeywell UDC2500
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Output Calibration
4/07 UDC2500 Universal Digital Controller Product Manual 155
Procedure
The procedure for calibrating the auxiliary output is listed in Table 6-4. The numeric
codes are also listed.
Make sure “LOCK” in the Tuning Set Up group is set to “NONE” (see Subsection 3.4).
Table 6-4 Auxiliary Output Calibration Procedure
(Numeric Code 50000)
Step Operation Press Result
1
Enter Calibration
Mode
SetupSetup
until you see
Upper Display = CAL ( - - - - )
Lower Display = AUXOUT (50000)
2
Calibrate 0 %
FunctionFunctionFunction
You will see:
Upper Display = A Value
Lower Display = ZROVAL (50001)
or
until the desired 0 % output is read on the milliammeter,
use the values shown below depending on the action of
your controller. Normally, this will be the setting that
produces 4 mA.
3
Calibrate 100 %
FunctionFunctionFunction
To store the 0 % value you will see:
Upper Display = A Value
Lower Display = SPNVAL (50002)
or
until the desired 100 % output is read on the milliammeter. .
Normally, this will be the setting that produces 20 mA.
4
FunctionFunctionFunction
The controller stores the span value.
Exit the
Calibration Mode
Lower
Display
Lower
Display
Lower
Display
To exit the calibration mode.

Table of Contents

Other manuals for Honeywell UDC2500

Related product manuals