EasyManua.ls Logo

Honeywell UDC2500 - Input 2 Calibration Procedure; Table 5-12 Set up Wiring Procedure for 0 to 2 Volts, 0 to 5 Volts, or 1 to 5 Volts - Input 2; Figure 5-9 Wiring Connections for 0 to 2 Volts, 0 to 5 Volts or 1 to 5 Volts Input - Input 2

Honeywell UDC2500
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Input Calibration
4/07 UDC2500 Universal Digital Controller Product Manual 147
0 to 2 Volts, 0 to 5 Volts, or 1 to 5 Volt Inputs – Input 2
Refer to Figure 5-9 and wire the controller according to the procedure given in Table
5-12.
Table 5-12 Set Up Wiring Procedure for 0 to 2 Volts, 0 to 5 Volts, or 1 to 5
Volts – Input 2
Step Action
1
Connect the copper leads from the calibrator to the Input #2 terminals as shown in
Figure 5-8.
2
Place voltage source at zero before switching on.
3
Do not switch voltage source ON/OFF while connected to the instrument.
Voltage
Source
26 +
27 -
Copper Leads
Equal Length
+
_
25 (no connection)
Voltage
Source
26 +
27 -
Copper Leads
Equal Length
+
_
25 (no connection)
Figure 5-9 Wiring Connections for 0 to 2 Volts, 0 to 5 Volts or 1 to 5 Volts
Input – Input 2
5.7 Input 2 Calibration Procedure
Preliminary Steps
Apply power and allow the controller to warm up for 30 minutes before you calibrate.
Please read Subsection
5.6 – before beginning the procedure.
Make sure you have LOCK set to NONE. See Subsection 3.4 - Tuning Set Up
Group.
Continued next page

Table of Contents

Other manuals for Honeywell UDC2500

Related product manuals