EasyManua.ls Logo

Honeywell UDC2500 - Table 5-13 Input 2 Calibration Procedure (Numeric Code 20000)

Honeywell UDC2500
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Input Calibration
148 UDC2500 Universal Digital Controller Product Manual 4/07
Procedure
The calibration procedure for Input #2 is listed in Table 5-13. The numeric codes are also
listed.
Table 5-13 Input 2 Calibration Procedure (Numeric Code 20000)
Step Operation Press Result
1
Enter Calibration
Mode
SetupSetup
until you see
Upper Display = CAL ( - - - - )
Lower Display = INPUT2 (20000)
FunctionFunctionFunction
You will see:
Upper Display = DIS ( 0 )
Lower Display = CALIN2 (20001)
or
You will see:
Upper Display = BEGN ( 1 )
Lower Display = CALIN2 (20001)
2
Calibrate 0 %
FunctionFunctionFunction
You will see:
Upper Display = APLY ( 2 )
Lower Display = IN2ZRO (20002)
Adjust your calibration device to an output signal equal
to the 0 % range value for your particular input sensor.
Wait 15 seconds, then go to the next step.
3
Calibrate 100 %
FunctionFunctionFunction
You will see:
Upper Display = APLY ( 2 )
Lower Display = IN2SPN (20003)
Adjust your calibration device to an output signal equal
to the 100 % range value for your particular input
sensor.
Wait 15 seconds, then go to the next step.
4
FunctionFunctionFunction
The controller stores the calibration constants.
Exit the
Calibration Mode
Lower
Display
Lower
Display
Lower
Display
To store the calibration constants and exit the calibration
mode.

Table of Contents

Other manuals for Honeywell UDC2500

Related product manuals