EasyManuals Logo

IAI EC-RR6*AH series User Manual

IAI EC-RR6*AH series
348 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #244 background imageLoading...
Page #244 background image
5.2 Maintenance information
5-3
5. Preventive/Predictive Maintenance
Overload warning level
Name Symbol Unit Input range
Default setting
at shipping
Overload warning
level
OLWL % 50 to 100
100
(disabled setting)
With the motor temperature rise value which triggers an overload alarm as 100%,
set the level to output an overload warning between 50 and 99%.
When the motor temperature rise value during operation exceeds the ratio set as
"overload warning level", the user will be notified with "Maintenance warning 3".
When the motor temperature rise value falls below the ratio set as
"overload warning level", "Maintenance warning 3" will be automatically cleared.
When "Maintenance warning 3" is output, investigate the cause of the increase
in load. Maintenance such as greasing or reviewing the operating conditions is
recommended.
Setting it to "100%" will disable this function.
Temperature change
under abnormal
conditions
Temperature change under normal conditions
Time
Warning output without stopping operation
Motor temperature
Stops operation
Overload alarm detection value
(100%)
Overload warning level set value
(50 to 99%)
Load increase
Alarm output
Warning
output
Perform
maintenance
Breakdowns and malfunctions can
be prevented by performing
maintenance after a warning is
output.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IAI EC-RR6*AH series and is the answer not in the manual?

IAI EC-RR6*AH series Specifications

General IconGeneral
BrandIAI
ModelEC-RR6*AH series
CategoryController
LanguageEnglish

Related product manuals