EasyManua.ls Logo

ITech IT-M3400 Series

ITech IT-M3400 Series
117 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Source Commands
Copyright © Itech Electronic Co., Ltd. 47
Default value
MIN
Example
CURR:UND:PROT 0.500
Query syntax
[SOURce:]CURRent:UNDer:PROTection[:LEVel]? [MINimum|MAXimum]
Returns
<NRf+>
[SOURce:]CURRent:UNDer:PROTection:DELay
<NRf+>
This command sets the under-current delay time of the power supply.
Syntax
[SOURce:]CURRent:UNDer:PROTection:DELay <NRf+>
Arguments
<0-10>
Default value
10.00S
Example
CURR:UND:PROT:DEL 10.000
Query syntax
[SOURce:]CURRent:UNDer:PROTection:DELay? [MINimum|MAXimum]
Returns
<NRf+>
[SOURce:]CURRent:UNDer:PROTection:STATe <bool>
This command enables or disables the under-current function.
Syntax
[SOURce:]CURRent:UNDer:PROTection:STATe <bool>
Arguments
<0|OFF|1|ON>

Table of Contents

Other manuals for ITech IT-M3400 Series