8 Trigger
8.6 Auto/Holdoff
having been below the threshold for at least the holdoff time. For negative slope, the
trigger event is generated if the signal characteristic crosses the threshold (with
negative slope) and then remains below the threshold for at least the holdoff time. In
either case, the trigger event is associated with the time the level was crossed
Remote Command
:TRIGger[:SEQuence]:HOLDoff:TYPE NORMal | ABOVe | BELow
:TRIGger[:SEQuence]:HOLDoff:TYPE?
Example
:TRIG:HOLD:TYPE NORM
Preset
Modes Setting
GSM/EDGE
Bluetooth
BELow
All others
NORMal
State Saved Saved in instrument state
1041 IQ Analyzer Mode User's &Programmer's Reference