4 System
4.6 Alignments
Perform Backup (without Flash) (Remote Command Only)
Invokes an alignment data backup operation to the provided location.
NOTE
Keysight recommends that the specified location should be external to the
instrument (USB or Mapped Network Drive).
Remote
Command
:CALibration:DATA:BACKup <filename>
Example
:CAL:DATA:BACK "F:\AlignDataBackup_N9020A_US00000001_2008140100.bkz"
Perform Restore (without Flash) (Remote Command Only)
Invokes an alignment data restore operation from the provided filename.
Remote
Command
:CALibration:DATA:RESTore <filename>
Example
:CAL:DATA:REST "F:\ AlignDataBackup_N9020A_US00000001_2008140100.bkz"