EasyManua.ls Logo

Keysight E6680A - Mass Storage Determine Removable Media Size; System:set (Remote Command Only)

Keysight E6680A
1175 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Save/Recall/Print
7.3 Save
Example
:MMEM:RMED:WPR? "F:"
Notes The return value is 1 if the device is write-protected, and 0 if the device is write-enabled
If the <partition> specified does not exist or is not a removable media device the error -252,
”Missing Media” is generated
Preset The return value depends on the SD card installed
7.3.13.13 Mass Storage Determine Removable Media size (Remote Query Only)
Queries a removable media device’s total memory size (not available memory size).
Remote
Command
:MMEMory:RMEDia:SIZE? <partition>
Example
:MMEM:RMED:SIZE? "F:"
Notes The return value is integer value in GBytes. Any device that is less than 1GB returns 0GB
If the <partition> specified does not exist or is not a removable media device, the error -
252,”Missing Media” is generated
7.3.13.14 :SYSTem:SET (Remote Command Only)
Obtains the state of the currently active mode in a form that can then be loaded
back into the instrument quickly.
Remote
Command
:SYSTem:SET <instrument state in IEEE Block>
:SYSTem:SET?
Notes The query returns current instrument state of the active mode in IEEE Block data format. The state is in
a machine-readable format only, as follows:
<syst set preamble><state block data>
Where:
<syst set preamble> is the format:
#NMMM
N = number of digits that comprise MMM
MMM = length in bytes of following data
<state block data> is machine readable state data
Example response: #42016<state data>
The state is recalled by sending the :SYST:SET? response data to the instrument. From example
above::SYST:SET #42016<state data>
IQ Analyzer Mode User's &Programmer's Reference 968