3 IQ Analyzer (Basic) Mode
3.1 Complex Spectrum Measurement
:CALC:SPEC:MARK3:STAT?
Preset
OFF
State Saved Saved in instrument state
Range
OFF|ON
Backwards
Compatibility
SCPI
Complex Spectrum:
:CALCulate:SPECtrum:MARKer[1]|2|…|12:STATe OFF | ON | 0 | 1
:CALCulate:SPECtrum:MARKer[1]|2|…|12:STATe?
Streaming:
:CALCulate:STReaming:MARKer[1]|2|…|12:STATe OFF | ON | 0 | 1
:CALCulate:STReaming:MARKer[1]|2|…|12:STATe?
Delta Marker (Reset Delta)
Pressing this control has exactly the same effect as pressing Delta in "Marker Mode"
on page 206. The selected marker becomes a DELTa Marker. If the selected marker
is already a DELTa marker, the reference marker is moved to the current position of
the selected marker, thus resetting the delta to zero.
Marker Settings Diagram
Lets you configure the Marker system using a visual utility.
All Markers Off
Enables you to turn off all markers.
Remote
Command
Complex Spectrum:
:CALCulate:SPECtrum:MARKer:AOFF
Streaming:
:CALCulate:STReaming:MARKer:AOFF
Example
:CALC:SPEC:MARK:AOFF
Couple Markers
When this function is ON, moving any marker causes an equal X-Axis movement of
every other marker that is not Off. By “equal X-Axis movement” we mean that we
preserve the difference between each marker’s X-Axis value (in the fundamental X-
Axis units of the trace that marker is on) and the X-Axis value of the marker being
moved (in the same fundamental X-Axis units).
207 IQ Analyzer Mode User's &Programmer's Reference