3 IQ Analyzer (Basic) Mode
3.1 Complex Spectrum Measurement
State Saved Saved in instrument state
Min/Max -10.0 s / 10.0s
Annotation <value> s bottom left of graph
Scale/Div
Lets you set the display X scale/division value. Only appears if the IQ Waveform
window is selected.
Remote
Command
Complex Spectrum:
:DISPlay:SPECtrum:VIEW[1]:WINDow2:TRACe:X[:SCALe]:PDIVision <time>
:DISPlay:SPECtrum:VIEW[1]:WINDow2:TRACe:X[:SCALe]:PDIVision?
Streaming:
:DISPlay:STReaming:VIEW[1]:WINDow2:TRACe:X[:SCALe]:PDIVision <time>
:DISPlay:STReaming:VIEW[1]:WINDow2:TRACe:X[:SCALe]:PDIVision?
Example
:DISP:SPEC:VIEW:WIND2:TRAC:X:PDIV 1e-9
:DISP:SPEC:VIEW:WIND2:TRAC:X:PDIV?
Couplings If X "Auto Scaling" on page 279 is ON, this value is automatically determined by the measurement
result. When you set a value manually, X Auto Scaling automatically changes to OFF
Preset 640.0 us !Burst Power
200.0 us !Waveform
State Saved Saved in instrument state
Min 1.00 ns
Max 1.0 s unless noted below:
320 s for Waveform
Ref Position
Lets you set the reference position for the X axis to: LEFT, CENTer or RIGHt. Only
appears if the IQ Waveform window is selected.
Remote
Command
Complex Spectrum:
:DISPlay:SPECtrum:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:RPOSition LEFT | CENTer |
RIGHt
:DISPlay:SPECtrum:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:RPOSition?
Streaming:
:DISPlay:STReaming:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:RPOSition LEFT | CENTer
| RIGHt
:DISPlay:STReaming:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:RPOSition?
IQ Analyzer Mode User's &Programmer's Reference 278