6 Input/Output
6.5 Corrections
Remote
Command
[:SENSe]:CCORection:CSET:COMMent "text"
[:SENSe]:CCORection:CSET:COMMent?
Example Firstly, select correction 4:
:CCOR:CSET:SEL 4
:CCOR:CSET:COMM "this is a comment"
Notes 60 chars max; may not fit on display if max chars used
Preset Unaffected by Preset. Set to empty by Restore Input/Output Defaults
State Saved Saved in instrument state
6.5.7.9 File
Shows the file from which the selected correction was loaded. If correction was
loaded with a SCPI command (see "Set Data (Remote Command Only)" on page
819) displays “(SCPI)”. If no correction is loaded, displays “(No correction loaded)”
Notes 60 chars max; may not fit on display if max chars used
State Saved Saved in instrument state
6.5.7.10 Freq Interpolation (Remote Command Only)
Controls how the correction values per-bucket are calculated. We interpolate
between frequencies in either the logarithmic or linear scale.
This setting is handled and stored individually per correction set.
VXT models M9410A/11A/15A/16A and M9410E/11E/15E/16E only support Linear
Interpolation.
See "Interpolation" on page 809 under Corrections.
Remote
Command
[:SENSe]:CCORrection:CSET:X:SPACing LINear | LOGarithmic
[:SENSe]:CCORrection:CSET:X:SPACing?
Example Firstly, select correction 4:
:CCOR:CSET:SEL 4
Set linear interpolation:
:CCOR:CSET:X:SPAC LIN
Preset Unaffected by Preset. Set to LINear by Restore Input/Output Defaults
State Saved Saved in instrument state
818 IQ Analyzer Mode User's &Programmer's Reference