6 Input/Output
6.1 RF Source
Note: the first PerArb trigger signal will be missed when the trigger delay is
negative.
External Trigger Polarity
Sets the polarity of the external trigger. When POSitive is selected, trigger event
happens on a rising edge of the external trigger in signal. When NEGative is
selected, trigger event happens on a falling edge of the external trigger in signal.
Active only if "Trigger Source" on page 675 is EXTernal1 or EXTernal2.
Remote Command
:SOURce:RADio:ARB:TRIGger[:SOURce]:EXTernal:SLOPe POSitive | NEGative
:SOURce:RADio:ARB:TRIGger[:SOURce]:EXTernal:SLOPe?
Example
:SOUR:RAD:ARB:TRIG:EXT:SLOP POS
:SOUR:RAD:ARB:TRIG:EXT:SLOP?
Dependencies Unavailable and grayed-out when "Trigger Source" on page 675 is not EXTernal1 or EXTernal2
Not available in E7760B
Preset
POSitive
Range
POSitive|NEGative
IQ Analyzer Mode User's &Programmer's Reference 679