3 IQ Analyzer (Basic) Mode
3.1 Complex Spectrum Measurement
Example
:SPEC:AVER 0
:SPEC:AVER?
Preset
ON
State Saved Yes
Range
ON|OFF
Average Mode
Lets you select the type of termination control used for the averaging function. This
determines the averaging action after the specified number of sweeps (average
count) is reached.
Available options are:
EXPonential - Each successive data acquisition after the average count is
reached, is exponentially weighted and combined with the existing average
REPeat - After reaching the average count, the averaging is reset, and a new
average is started
Remote
Command
Complex Spectrum:
[:SENSe]:SPECtrum:AVERage:TCONtrol EXPonential | REPeat
[:SENSe]:SPECtrum:AVERage:TCONtrol?
Streaming:
[:SENSe]:STReaming:AVERage:TCONtrol EXPonential | REPeat
[:SENSe]:STReaming:AVERage:TCONtrol?
Example
:SPEC:AVER:TCON REP
:SPEC:AVER:TCON?
Preset
EXPonential
State Saved Yes
Range
EXPonential|REPeat
Average Type
Lets you select the averaging type:
Option Parameter Behavior
Pwr Avg (RMS)
RMS
The power is averaged, providing the rms of the
voltage.LOG
Log Pwr Avg
(Video)
LOG
The log of the power is averaged