6 Input/Output
6.10 Calibrator Control
Preset Unaffected by Mode Preset. Set to 0Hz by Restore Input/Output Defaults or Restore System
Defaults->All
Min/Max Dependent on the selected calibrator
6.10.6 Calibrator Reference
Determines the frequency reference type used by the RCal module of the currently
selected Cal Group
Remote Command
:SYSTem:CALibration:TUNE:REFerence INTernal | EXTernal
:SYSTem:CALibration:TUNE:REFerence?
Example Set the calibrator frequency reference to Internal:
:SYSTem:CALibration:TUNE:REFerence INTERNAL
Dependencies Only displayed when an RCal module is the selected calibrator
Preset Unaffected by Mode Preset. Set to preset value by Restore Input/Output Defaults or Restore System
Defaults->All
Range
INTernal|EXTernal
904 IQ Analyzer Mode User's &Programmer's Reference