3 IQ Analyzer (Basic) Mode
3.1 Complex Spectrum Measurement
[:SENSe]:SPECtrum:LO:MIXMode?
Streaming:
[:SENSe]:STReaming:LO:MIXMode NORMal | ALTernate
[:SENSe]:STReaming:LO:MIXMode?
Example
:SPEC:LO:MIXM NORM
:SPEC:LO:MIXM?
Dependencies Grayed-out when the RF Input is set to something other than RF (such as External Mixer). If you press
the grayed-out control, a warning message “Feature only available with Signal Input RF” is shown
If you try to set Mixing Mode via SCPI when disabled, a message -221, "Settings conflict; Feature only
available with signal input RF" is returned
Preset
NORMal
State Saved Yes
Range
NORMal | ALTernate
ADC Correction (Backwards Compatibility only)
Performs ADC calibration.
Preset 1
Backwards Compatibility
SCPI
Not supported in STReaming measurement
[:SENSe]:SPECtrum:WBIF:ADCCorrect ON | OFF | 1 | 0
[:SENSe]:SPECtrum:WBIF:ADCCorrect?
IF Flatness (Backwards Compatibility only)
Turns on and off IF flatness corrections.
Preset 1
Backwards Compatibility
SCPI
Not supported in STReaming measurement
[:SENSe]:SPECtrum:WBIF:FLATness OFF | ON | 0 | 1
[:SENSe]:SPECtrum:WBIF:FLATness?
Power Reference Plane
Allows you to increase the power by 3dB for Baseband I+jQ measurements. Options
are:
Menu SCPI
Baseband
BASeband
RF
RF
265 IQ Analyzer Mode User's &Programmer's Reference