6 Input/Output
6.1 RF Source
Time or Count <time/int> - specifies time duration in seconds or play count of the ARB file associated
with the step
If input parameter number exceeds the step number defined by "Number of Steps" on page 621, then
an error is generated, and only those parameters whose index number falls within number of steps will
be updated
If current "Step Duration" on page 631 is “Continuous”, then error -221, "Settings conflict; Cannot
accept time or count input when step duration type is Continuous on step #" is generated
Dependencies The range is 1 to 1000, which is determined by the number of steps you have configured. For details
see "Number of Steps" on page 621
Step Configuration of Output Trigger parameter list (Remote Command Only)
Configures the Output Trigger parameter array of the whole List Sequencer at one
time. The number of arrays is same as step number defined in "Number of Steps" on
page 621. As a step is set up, the value entered runs through several levels of
validation.
Remote Command
:SOURce:LIST:SETup:OUTPut:TRIGger <bool>, <bool>, <bool>, …
:SOURce:LIST:SETup:OUTPut:TRIGger?
Example
:SOUR:LIST:SET:OUTP:TRIG ON,OFF,ON
:SOUR:LIST:SET:OUTP:TRIG?
Notes Sets up or queries the parameter array of whole list sequence
Output Trigger <Boolean> - specifies the output trigger for the step. Options are:
ON|OFF|1|0
If input parameter number exceeds the step number defined by "Number of Steps" on page 621, then
error -221,"Settings conflict; The number of input parameters is too large and is truncated to current
list step number" is generated, and only those parameters whose index number falls within legal step
number are updated
Dependencies The range is 1 to 1000, which is determined by the number of steps you have configured. For details
see "Number of Steps" on page 621
Clear List (Remote Command Only)
The SCPI equivalent of the Clear List UI feature described in "Clear List" on page
622.
Remote Command
:SOURce:LIST:SETup:CLEar
Example
:SOUR:LIST:SETup:CLE
Dependencies Not available in E7760B