7 Save/Recall/Print
7.3 Save
Column Value
9 Trace data containing the shape of the FFT window
10 Trace data of the phase of the FFT versus frequency
11 Linear spectrum trace data in Volts RMS
12 Linear spectrum trace data in Volts RMS
13 A set of 13 results, as listed in "Complex Spectrum Measurement" on page 143
14 Current delta magnitude trace in dB
15 Current delta phase trace in radians
16 Averaged delta magnitude trace in dB
17 Averaged delta phase trace in radians
Example Data Block
The following is an example of the Data Block of a Meas Results file, as it would
appear when imported into Microsoft Excel, or similar spreadsheet application.
MeasResult1 MeasResult2 MeasResult3 MeasResult4 … MeasResult17
-22.19829559 -14.78834 0.046175 -104.228 … -94.904
63145289.72 -14.79132 -0.03447 -98.5835 -95.1532
205 -14.78322 0.056872 -94.8855 -95.2769
60047250.51 -14.78362 -0.00914 -92.0313 -95.9835
39215.68627 -14.77991 0.055141 -91.0622 -95.5153
237 -14.77921 0.016839 -89.0501 -95.0541
0 -14.7752 0.043001 -87.94 -95.7244
1.00E-07 -14.77521 0.038403 -87.2431 -95.0759
1 -14.77495 0.024597 -88.0133 -94.8669
2.36E-05 -14.7738 0.052169 -90.5852 -94.4348
25 -14.77691 0.003863 -104.132 -95.0335
Rows continue until all data is displayed
IQ Data (Normal)
Saves IQ data with the number of points defined by the formula below.
Remote Command
:MMEMory:STORe:SPECtrum:IQData "File name"
Example
:MMEM:STOR:SPEC:IQD "myResults.txt"
Saves IQ data from Complex Spectrum measurement into myResults.txt in the
BASIC/data/SPEC/results directory on the data drive
Dependencies Only available in IQ Analyzer (BASIC) Mode
949 IQ Analyzer Mode User's &Programmer's Reference