3 IQ Analyzer (Basic) Mode
3.1 Complex Spectrum Measurement
“MyData_0001.binf”
“MyData_0002.binf”
and so on.
Not available in the STReaming measurement.
3.1.10 Trace
Lets you control the display and storage of trace data for the available traces.
3.1.10.1 Settings
Lets you configure and execute Normalize functions to display one trace relative to a
reference trace.
Spectrum Trace Display
Toggles the instantaneous trace display ON or OFF.
This is a display-only feature. The actual Spectrum trace still exists, even when the
Spectrum Trace State is OFF, but it is just not visible. Therefore, it is possible to set
Spectrum Trace State to OFFand put a marker on Spectrum trace.
Remote Command Complex Spectrum:
:TRACe:SPECtrum:STRace[:STATe] OFF | ON | 0 | 1
:TRACe:SPECtrum:STRace[:STATe]?
Streaming:
:TRACe:STReaming:STRace[:STATe] OFF | ON | 0 | 1
:TRACe:STReaming:STRace[:STATe]?
Example
:TRAC:SPEC:STR OFF
:TRAC:SPEC:STR?
Preset
ON
State Saved Yes
Range
OFF|ON
IQ Analyzer Mode User's &Programmer's Reference 282