6 Input/Output
6.1 RF Source
Select PXI Line
Controls which PXI_TRIG[0..7] backplane line is used for the trigger source.
Only appears in modular analyzer products.
Remote Command
:SOURce:RADio:ARB:TRIGger[:SOURce]:PXI:LINE <line>
:SOURce:RADio:ARB:TRIGger[:SOURce]:PXI:LINE?
Example
:SOUR:RAD:ARB:TRIG:PXI:LINE 2
Dependencies Unavailable and grayed-out when "Trigger Source" on page 675 is not set to PXI
Not available in E7760B
Preset 0
State Saved Saved in instrument state
Range [0,7]
PXI Trigger Delay
Lets you toggle the state and value of PXI trigger delay. The value you enter sets a
delay time between when an PXI trigger is received and when it is applied to the
waveform.
Only active if "Trigger Source" on page 675 is PXI.
Remote Command
:SOURce:RADio:ARB:TRIGger[:SOURce]:PXI:DELay <time>
:SOURce:RADio:ARB:TRIGger[:SOURce]:PXI:DELay?
Example
:SOUR:RAD:ARB:TRIG:PXI:DEL 100ns
:SOUR:RAD:ARB:TRIG:PXI:DEL?
Notes PXI trigger delay time set by users will be rounded to the nearest integer multiple of the resolution
Dependencies Unavailable and grayed-out when "Trigger Source" on page 675 is not PXI
Not available in E7760B
Preset 1 ms
Min VXT models M9410A/11A/15A/16A and M9410E/11E/15E/16E: -10 s
Max
Instrument/Condition Value Derivation
VXT models M9410A/11A/15A/16A 11.45324612 s 2.666667ns * (2^32-1)
M9410E/11E/15E/16E 11.45324612 s 2.666667ns * (2^32-1)
Continuous Trigger + Run” trigger 11.45324612 s 2.666667ns * (2^32-1)
Other trigger conditions 17.17986918 s 4 ns *(2^32-1)
All Others 8.589934588 s 4ns * (2^31 1)