EasyManua.ls Logo

Keysight E6680A - Mixer Lvl Offset; Range (Baseband Input Models)

Keysight E6680A
1175 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 IQ Analyzer (Basic) Mode
3.1 Complex Spectrum Measurement
Example
:POW:RANG:PAR 12 dB
Notes In some Modes, this parameter is read-only; meaning the value will appear on the control and query
via SCPI, but is not changeable. In such applications the control is grayed-out. Attempts to change the
value via SCPI are ignored, but no error message is generated
Dependencies Does not appear in Spectrum Analyzer Mode
Preset
VXT Models M9410A/11A 0dB
All Others 10dB
State Saved Saved in instrument state
Min 0dB
Max
VXT Models M9410A/11A 50dB
All Others 20dB
Mixer Lvl Offset
This is an advanced setting to adjust target Range at the input mixer, which in turn
affects the signal level in the instrument’s IF. This setting can be used when
additional optimization is needed after setting "Peak-to-Average Ratio" on page
314. Positive values of offset optimize noise performance over distortion, negative
values optimize distortion performance over noise.
Remote Command
[:SENSe]:POWer[:RF]:RANGe:MIXer:OFFSet <real>
[:SENSe]:POWer[:RF]:RANGe:MIXer:OFFSet?
Example
:POW:RANG:MIX:OFFS -5 dB
Preset 0dB
State Saved Saved in instrument state
Min
VXT Models M9410A/11A -34dB
All Others -35dB
Max 30dB
3.1.3.4 Range (Baseband Input models)
Only available when Option BBA is present (I/Q Baseband Inputs), the current
measurement supports option BBA, and I/Q is the selected input. In these cases,
replaces the Attenuation tab.
163 IQ Analyzer Mode User's &Programmer's Reference