3 IQ Analyzer (Basic) Mode
3.2 IQ Waveform Measurement
RF Envelope
Remote
Command
:DISPlay:WAVeform:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RLEVel <amptd>
:DISPlay:WAVeform:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RLEVel?
Example
:DISP:WAV:VIEW:WIND:TRAC:Y:RLEV -50 dBm
Couplings When "Auto Scaling" on page 292 is ON (default), this value is automatically determined by the
measurement result. When you set a value manually, Auto Scaling changes to OFF
Preset 10.00 dBm
State Saved Saved in instrument state
Min/Max –/+250.00 dBm
Annotation Ref <value> top of graph
IQ Waveform
Remote
Command
:DISPlay:WAVeform:VIEW2:WINDow[1]:TRACe:Y[:SCALe]:RLEVel <voltage>
:DISPlay:WAVeform:VIEW2:WINDow[1]:TRACe:Y[:SCALe]:RLEVel?
Example
:DISP:WAV:VIEW2:WIND:TRAC:Y:RLEV 25 V
Couplings When "Auto Scaling" on page 292 is On (default), this value is automatically determined by the
measurement result. When you set a value manually, Auto Scaling changes to Off
Preset 0 V
State Saved Saved in instrument state
Min/Max –/+250.00 V
Annotation Ref <value> top of graph
Scale/Div
Enables you to set the units per division of vertical scale in the logarithmic display.
However, since "Auto Scaling" on page 292 defaults to ON, this value is
automatically determined by the measurement result. When you set a value
manually, Auto Scaling automatically changes to OFF.
The SCPI command and default parameters are dependent on whether the View is
"RF Envelope" on page 285 or "I/Q Waveform" on page 286.
RF Envelope
Remote
:DISPlay:WAVeform:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:PDIVision <rel_ampl>
289 IQ Analyzer Mode User's &Programmer's Reference