3 IQ Analyzer (Basic) Mode
3.2 IQ Waveform Measurement
Peak Search Range
Lets you specify the range of trace to perform peak search actions specified in the
Peak Search menu.
Option SCPI Behavior
Full
FULL
Peak Search actions will be performed on the entire trace
Manual
MANual
Specifies the range of the trace to which Peak Search actions will
apply
Remote
Command
:CALCulate:WAVeform:MARKer:PEAK:SEARch:RANGe FULL | MANual
:CALCulate:WAVeform:MARKer:PEAK:SEARch:RANGe FULL | MANual
Example
:CALC:WAV:MARK:PEAK:SEAR:RANG FULL
Preset
FULL
State Saved Saved in instrument state
Range
FULL | MANual
Peak Search Range Start
Specifies the start of the range, in seconds, to which Peak Search actions are
applied. Displays as a green vertical line, with the label Pk Search Start on the left,
as shown below.
Remote Command
:CALCulate:WAVeform:MARKer:PEAK:SEARch:RANGe:STARt <time>
:CALCulate:WAVeform:MARKer:PEAK:SEARch:RANGe:STARt?
Example
:CALC:WAV:MARK:PEAK:SEAR:RANG:STAR 0.001
Dependencies Disabled when Peak Search Range is set to Full.
Preset 0.0 ms
Min/Max 0.0/+Meas Time
Peak Search Range Stop
Specifies the end of the range, in seconds, to which Peak Search actions are
applied. Displays as a green vertical line, with the label Pk Search Stop on the left,
361 IQ Analyzer Mode User's &Programmer's Reference