3 IQ Analyzer (Basic) Mode
3.2 IQ Waveform Measurement
gain setting, but you must ensure that your measurement conditions will not
overload the IF (in the HIGH gain range), that your signals are well above the noise
floor (for the LOW gain range), and that the signals are not changing.
When Digital Bus Out (under the Input/Output menu) is ON, the IF Gain State
AUTOrange selection is not allowed. Thus, in this case IF Gain State will be set to
LOW.
This only applies to the RF input. It does not apply to baseband I/Q input.
Remote Command
[:SENSe]:WAVeform:IF:GAIN[:STATe] AUTOrange | LOW | HIGH | OTHer
[:SENSe]:WAVeform:IF:GAIN[:STATe]?
Example
:WAV:IF:GAIN HIGH
:WAV:IF:GAIN?
Notes Only applies to the RF input. Does not apply to baseband I/Q input
Dependencies If you try to select AUTOrange via SCPI while Digital Bus Out is ON, an error message -224, “Illegal
parameter value; "IF Gain Autorange not allowed when Digital Bus Out is on" is displayed
If you try to select AUTOrange via the front panel while Digital Bus Out is ON, an error message -221
“Settings conflict; "IF Gain Autorange not allowed when Digital Bus Out is ON" is displayed
Other IF Gain is available only in models with DIF40
Couplings As for most parameters that have an AUTO state, "Auto Couple" on page 374 sets it to AUTOrange,
which then selects LOW or HIGH depending on the IF Path. Setting any specific value (AUTOrange,
LOW, HIGH, or OTHer) sets the AUTO state to OFF
Preset
LOW
State Saved Saved in instrument state
Range Autorange (Slower Follows Signals) | Low (Best for Large Signals) | High (Best Noise Level) | Other
(Explicit)
Auto Function
Remote Command
[:SENSe]:WAVeform:IF:GAIN:AUTO[:STATe] ON | OFF | 1 | 0
[:SENSe]:WAVeform:IF:GAIN:AUTO[:STATe]?
Example
:WAV:IF:GAIN:AUTO ON
:WAV:IF:GAIN:AUTO?
Notes Activates the auto rules for IF Gain
Preset
ON
Range
OFF|ON
IF Gain Offset
Sets IF Gain Offset for the 40 MHz, 140 MHz, 160 MHz IF Paths in 1dB steps from
the minimum gain available to the maximum. Increasing the gain can increase the
393 IQ Analyzer Mode User's &Programmer's Reference