3 IQ Analyzer (Basic) Mode
3.3 Streaming Measurement
3.3.2 Windows
The following windows are available in the Streaming measurement.
Spectrum Window
The Spectrum window displays amplitude versus frequency information.
RF Envelope Window
The RF Envelope window displays amplitude versus time information.
Status Window
The Status window displays statistics for the streaming session in progress.
3.3.3 Meas Setup
In the Streaming measurement, the Meas Setup panel has the following additional
tabs, relative to the "Complex Spectrum Measurement" on page 143.
3.3.3.1 Streaming
Contains controls to set up streaming, and to start or stop streaming.
Dependencies Only appears if you have the proper option installed in your instrument
Streaming Duration
Specifies the duration of the streaming session with FINite Streaming. Disabled
when "Streaming Duration Type (Auto Function)" on page 417 is set to INFinite.
Remote Command
[:SENSe]:STReaming:ACQuisition:DURation <time>
[:SENSe]:STReaming:ACQuisition:DURation?
Example
:STR:ACQ:DUR 10
:STR:ACQ:DUR?
Notes You cannot change this value while streaming is in progress. If Streaming Duration Type is set to
FINite and you try to change the value, it will generate the error message:
IQ Analyzer Mode User's &Programmer's Reference 416