9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Questionable Condition
Returns the decimal value of the sum of the bits in the Questionable Condition
register.
NOTE
The data in this register is continuously updated and reflects current conditions.
Remote Command
:STATus:QUEStionable:CONDition?
Example
:STAT:QUES:COND?
Preset 0
Status Bits/OPC depend-
encies
Sequential command
Questionable Enable
Determines which bits in the "Questionable Event Query" on page 1119 Register will
set the Questionable Status Summary bit (bit3) in the "Status Byte Register" on
page 1104.
The variable <integer> is the sum of the decimal values of the bits you want to
enable.
NOTE
The preset condition is all bits in this enable register set to 0. To report any
Questionable Events to the Status Byte Register, one or more bits need to be set
to 1. The "Standard Event Status Register" on page 1108 should be queried after
each measurement to check the Questionable Status Summary (bit 3). If it is
equal to 1, a condition during the test may have made the test results invalid. If it
is equal to 0, this indicates that no hardware problem or measurement problem
was detected by the analyzer.
Remote
Command
:STATus:QUEStionable:ENABle <integer>
:STATus:QUEStionable:ENABle?
:STATus:OPERation:ENABle <integer>
:STATus:OPERation:ENABle?
Example
:STAT:QUES:ENAB 16
Sets the register so that questionable temperature events will be reported to the Status Byte Register
Preset 0
Min 0
Max 32767
Status Bits/OPC Sequential command