9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:PTRansition <integer>
:STATus:QUEStionable:PTRansition?
Example
:STAT:QUES:PTR 16
Temperature summary ‘questionable asserted’ will be reported to the Status Byte Register
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
9.4.6.6 Questionable Power Register
Monitors power-related conditions within the instrument and summarizes them in
bit 3 of the "Questionable Register" on page 1117.
Bit Condition Operation
0 RPP tripped (not currently in use)
1 Source Unleveled The built-in source is not properly leveled
2 Source LO
Unleveled
(not currently in use)
3 LO Unleveled (not currently in use)
6 Input Port
Overload
A power overload condition exists at an input port
9 Full Bypass
Enabled
Frontend circuitry is bypassed, use caution to protect the mixer
10 Input Caution Input circuitry is configured such that care is required to prevent
damage
Filter Registers
IQ Analyzer Mode User's &Programmer's Reference 1120