10 Hardware-Accelerated Fast Power Measurement (Remote Command Only)
10.3 Define Fast Power Measurement Query (Remote Command Only)
10.3 Define Fast Power Measurement Query (Remote Command
Only)
Retrieves a list of all defined parameters in an ASCII string format
The following is an example of returned results:
"DCCoupled=False,ElecAttBypass=True,ElecAttenuation=0,IFGain=0,MechAttenuation=
0,PreAmpMode=Off,PreSelectorOffset=0,UsePreSelector=False,ExternalReferenceFreq
uency=10000000,FrequencyReferenceSource=AutoExternalFrequencyReference,IFType=B
40M,LOMode=SLW,SignalInput=FpMainRf,AcquisitionTime=0.001,CenterFrequency=10000
00000,ResolutionBW=0,ResolutionBWMode=BestSpeed,DetectorType=RmsAverage,Bandwid
th=[1000000],OffsetFrequency=[0],Function=[BandPower],FilterType=
[IBW],FilterAlpha=[0.22],OccupiedBandwidthPercent=[0.99],XdBBandwidth=
[3.01],DoNoiseCorrection=False,DoSpurSuppression=False,MeasurementMethod=Hardwa
reFFT,IncludePowerSpectrum=False,TriggerDelay=0,TriggerLevel=1.2,TriggerSlope=P
ositive,TriggerSource=Free,TriggerTimeout=1
,Trigger1Output=Off,Trigger1OutputPolarity=Positive,Trigger2Output=Off,Trigger2
OutputPolarity=Positive"
Remote Command
:CALCulate:FPOWer:POWer[1,2,…,999]:DEFine?
Example
:CALC:FPOW:POW1:DEF?
Notes Retrieves a list of all defined parameters in an ASCII format
IQ Analyzer Mode User's &Programmer's Reference 1168