Marker Mode
These controls are available when the “Single Channel with Marker” or “Dual Channel
with Marker” mode is selected in the Output tab.
Ch. 3 (Marker 1)
Marker 1 is output on Channel 3. Signaling the beginning of each segment
may be activated (Segment selection) and deactivated (None selection).
Ch. 4 (Marker 2)
Marker 2 is output on Channel 4. Signaling the beginning of each segment
may be activated (Segment selection) and deactivated (None selection).
Scaling Section
DAC Max
Standard waveforms may occupy a limited range of the DAC’s full scale. This
parameter sets the maximum level. If set to a lower level than DAC Min, this
will be automatically set to the same level. Acceptable range for this
parameter is -1/+1, being the full dynamic range of the instrument’s DAC.
DAC Min
Standard waveforms may occupy a limited range of the DAC’s full scale. This
parameter sets the minimum level. If set to a higher level than DAC Max, this
will be automatically set to the same level. Acceptable range for this
parameter is -1/+1, being the full dynamic range of the instrument’s DAC.