2G0947 - 31JAN08 3-79
Adjustments and Replacements
SUCTION CUP
Prerequisites: Remove the “PICKUP AY - UPPER” or “PICKUP AY - LOWER”.
To Remove:
This procedure is used for all PICKUP AY SUCTION CUPS.
1. Remove each SUCTION CUP as follows:
a. Insert a small flat-blade screwdriver into the slot at the center of each
SUCTION CUP and turn counterclockwise to remove the SUCTION
CUP and HUB.
b. Remove the HUB from the SUCTION CUP.
To Install: 1. Reverse the steps in the removal procedure.
HUB
SUCTION CUP