2G0947 - 31JAN08 2-13
Installation Instructions
Doing the Configuration
You must restart the IMAGER from the SERVICE TOOL in order to
permanently save configuration information.
Setting the Language
The language is set by the customer in the WEB PORTAL.
Checking the Values for the DICOM Source Communications
1. Connect the LAPTOP COMPUTER to the IMAGER.
2. Select Configuration>DICOM SCP>Communications.
3. Check:
• "Port Number" is 5040
• "Association Limit" is 12
4. Are the values correct?
Yes No
Continue with “Setting the
IMAGER to be a Destination”.
a. Click [Modify].
b. Type the correct values.
c. Click [Save].
d. Continue with “Setting the
IMAGER to be a Destination”.