3-54 2G0947 - 31JAN08
Adjustments and Replacements
MOTOR, NIP POSITION DRIVE (M2)
Prerequisites: Remove:
“BACK PANEL”
“RIGHT PANEL”
To Remove:
1. Disconnect:
MOTOR CONNECTOR
4 CIRCUIT BREAKER CONNECTORS
2. Loosen the TENSIONER SCREW.
3. Disengage the BELT for the PULLEY on the NIP POSITION MOTOR.
4. Remove:
4 SCREWS
NIP POSITION MOTOR
To Install: 1. Reverse the steps in the removal procedure.
NIP POSITION
MOTOR
MOTOR
CONNECTOR
Back of IMAGER
4 SCREWS
TENSIONER
SCREW
BELT
4 CIRCUIT BREAKER
CONNECTORS