2G0947 - 31JAN08 5-15
Diagnostics
Condition Code: 01-010
Description
A subsystem (xxxx) that was started by the Configuration and Control
Subsystem terminated abnormally.
System Impact
The subsystem will be temporarily unavailable until it is restarted by the
Configuration and Control Subsystem. Depending on the subsystem, it may
cause associations to be rejected, or printing to be delayed.
User Action
None
FE Action
None, this is a log notification only.
Condition Code: 04-000
Description
A general, non-specific error has occurred in the DICOM Input Print Service
Class Provider (SCP) Subsystem.
System Impact
Unknown.
User Action
None
FE Action
None, this is a log notification only.
Web Portal Message Log Message DICOM Message Severity
None Subsystem xxxx terminated. NORMAL 3-Notify
Web Portal Message Log Message DICOM Message Severity
None General Error in DICOM Input
Print SCP Subsystem.
NORMAL 3-Notify