5-52 2G0947 - 31JAN08
Diagnostics
Condition Code: 20-706
Description
This auxiliary code indicates that a user-initiated system shutdown is complete.
System Impact
The printer is unable to print films.
User Action
1. Shut the IMAGER power OFF.
FE Action
1. De-energize the IMAGER.
Condition Code: 20-913
Description
This notification occurs when the wrong number of pixels is transferred between
the DRE and the DPB in the allotted timeframe. The faulty image will be
printed. The job will be failed, temporarily placed in the undelivered queue, and
rescheduled for delivery. (Note: Error 20-915 is generated when this problem is
detected on the 3rd successive film.)
System Impact
None - the system will function normally.
User Action
None
FE Action
None, this is a log notification only.
Web Portal Message Log Message DICOM Message Severity
None (none) (Not Applicable) 4-Auxiliary
Web Portal Message Log Message DICOM Message Severity
None Image transfer failure. Wrong
number of bytes transferred in
the allotted timeframe.
NORMAL 3-Notify