5-78 2G0947 - 31JAN08
Diagnostics
Condition Code: 25-922
Description
When the RF Tag reader is told to perform diagnostics and returns a diagnostics 
failure, the Machine Control System (MCS) commands a retry. This error is 
indicated if the MICROCONTROLLER still indicates a diagnostics failure after 
3 attempts. The error clears only when power is cycled. 
System Impact
The printer is unable to print films. The spooler is still able to accept and queue 
jobs but new films will not be initiated. Films downstream of the affected 
subsystem will be completed. All other films will be reprinted. 
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service. 
FE Action
Condition Code: 25-931
Description
This error occurs when the Machine Control System (MCS) tries to 
communicate with the MICROCONTROLLER but receives no response to 5 
attempts. The error clears only when power is cycled. 
System Impact
The printer is unable to print films. The spooler is still able to accept and queue 
jobs but new films will not be initiated. Films downstream of the affected 
subsystem will be completed. All other films will be reprinted. 
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service. 
FE Action
Web Portal Message Log Message DICOM Message Severity
RF Tag: Internal Diagnostic 
Failure 
RF Tag reader failed self 
diagnostics
BAD SUPPLY MGZ  0-Fatal 
Web Portal Message Log Message DICOM Message Severity
RF Tag: Internal 
Communications Failure 
MCS failed to communicate 
with RF Tag micro after 
multiple attempts
ELEC DOWN  0-Fatal