5-98 2G0947 - 31JAN08
Diagnostics
Condition Code: 29-154
Description
This error occurs when the MICROCONTROLLER reports that it has detected
an error while attempting to read or write from its NVRAM. The error clears
only when power is cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Web Portal Message Log Message DICOM Message Severity
Densitometer: Internal
Communications Failure
Densitometer microcontroller
failed to read/write to NVRAM
ELEC DOWN 0-Fatal