Condition Code: 29-926
Description
During a calibration print and during DENSITOMETER diagnostics, a reading
from the FIDUCIAL SENSOR S11) is taken with the light source turned off. The
error is indicated if the MICROCONTROLLER reports the reading out of range.
The error clears only when power is cycled.
System Impact
The printer is unable to print films. The spooler is still able to accept and queue
jobs but new films will not be initiated. Films downstream of the affected
subsystem will be completed. All other films will be reprinted.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Web Portal Message Log Message DICOM Message Severity
Densitometer: Internal
Diagnostic Failure
Densitometer fiducial sensor
offset reading out of range with
light source off
ELEC DOWN 0-Fatal