5-102 2G0947 - 31JAN08
Diagnostics
Condition Code: 29-931
Description
This error occurs when the Machine Control System (MCS) tries to 
communicate with the MICROCONTROLLER but receives no response to 5 
attempts. The error clears only when power is cycled. 
System Impact
The printer is unable to print films. The spooler is still able to accept and queue 
jobs but new films will not be initiated. Films downstream of the affected 
subsystem will be completed. All other films will be reprinted. 
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service. 
FE Action
Condition Code: 29-932
Description
This notification occurs when the Machine Control System (MCS) tries to 
communicate with the MICROCONTROLLER but receives no response. (Note: 
Error 29-931 is generated when this problem is detected on the 5th successive 
failed communication attempt.) 
System Impact
None - the system will function normally. 
User Action
None 
FE Action
None, this is a log notification only.
Web Portal Message Log Message DICOM Message Severity
Densitometer: Internal 
Communications Failure 
MCS failed to communicate 
with Densitometer after 
multiple attempts
ELEC DOWN  0-Fatal 
Web Portal Message Log Message DICOM Message Severity
None  MCS failed to communicate 
with Densitometer
NORMAL 3-Notify