5-104 2G0947 - 31JAN08
Diagnostics
Condition Code: 29-947
Description
During a calibration print and during DENSITOMETER diagnostics, a reading
from the FIDUCIAL SENSOR (S11) is taken with the light source turned on.
This notification occurs when the reading is outside of the optimal range, i.e.
approaching the range limit. (Note: Error 29-927 is generated when this reading
is over the range limit.)
System Impact
None - the system will function normally.
User Action
None
FE Action
None, this is a log notification only.
Condition Code: 36-931
Description
This error occurs when the Machine Control System (MCS) tries to
communicate with the MICROCONTROLLER but receives no response to 5
attempts. The error clears only when power is cycled.
System Impact
The LOCAL PANEL User Interface will not display correct data.
User Action
1. Shut down the IMAGER, then start the IMAGER.
2. If the error persists, call for service.
FE Action
Web Portal Message Log Message DICOM Message Severity
None Densitometer fiducial sensor
reference reading outside of
optimal range with light source
on
NORMAL 3-Notify
Web Portal Message Log Message DICOM Message Severity
Local Panel: Internal
Communications Failure
MCS failed to communicate
with Local Panel after multiple
attempts
ELEC DOWN 0-Fatal