2G0947 - 31JAN08 5-109
Diagnostics
Film Jam - Error Code 2x-126
1. Remove the FILM TRAYS.
2. Remove any misplaced films from Area 1. This area can be accessed 
through the FILM TRAY slots. 
3. If Area 1 did not contain misplaced films, take the FILM TRAYS to a 
dark room and remove the FILM SAVER. Remove any misplaced films 
and replace the FILM SAVER.
4. If misplaced films were not found in the TRAY, open and close the TOP 
COVER. Wait 2 minutes for misplaced films to emerge from the 
IMAGER. 
If misplaced films were never located and this error recurs on the next 
film, continue with Step 5.
5. Open the TOP COVER.
6. Loosen 2 Captice THUMB SCREWS.
7. Remove the LEFT PANEL.
TOP COVER
2 Captive
THUMB
SCREWS
LEFT PANEL